Universal Library User's Guide Temperature Input Boards - DAS-TC Series
184
DAS-TC Series
Temperature Input
Temperature input functions and methods supported
UL: cbTIn(), cbTInScan()
UL for .NET: TIn(), TInScan()
Temperature input argument values
Options NOFILTER
Scale CELSIUS, FAHRENHEIT, KELVIN
HighChan 0 to 15
Hardware considerations
Pacing input
The rate of measurement is fixed at approximately 25 samples per second.
Selecting thermocouples
J, K, E, T, R, S, N, or B type thermocouples may be selected using InstaCal.
Open thermocouples
When using cbTInScan() or TInScan() with the DAS-TC, an open thermocouple error (OPENCONNECTION) on
any of the channels will cause all data to be returned as –9999.0. This is a hardware limitation. If your
application requires isolating channels with defective thermocouples attached and returning valid data for the
remainder of the channels, use the cbTIn() function or TIn() method instead.
To read the voltage input of the thermocouple, select VOLTS for the Scale parameter in cbTIn() and
cbTInScan(), or TIn() and TInScan().